|
Active Thermography: An Overview
of
Methods and Their Applications in Use Today |
|
|

Michael R. Sharlon |
Michael R.
Sharlon, President
Thermasearch, Inc.
1215 Sturgis
Road
Conway, AR 72034
Tel: 501-513-9901
|
|
Abstract
Currently, there are several
forms of active thermography. This paper will introduce
thermographers to the different methods of active thermography
and will give a brief overview of each method. Some types
of active thermography discussed will include reflective
or transmissive heating as applied to pulsed, pulsed phase,
lock-in, vibro, thermoelastic stress analysis and thermomechanical
couplings in solids thermography. Also covered will be current
and possible future uses for active thermography.
Introduction
Active Thermography (AT) is
defined as applying a stimulus to a target to cause the target
to heat or cool in such a way as to allow characteristics
of the target to be observed when viewed by thermal imagery.
These observed characteristics may be flaws sought in Non
Destructive Testing (NDT) or norms sought in quality control.
What separates active from passive thermography
is the intentional application of heating, chilling or stress
(active) that results in a temperature rise or fall in a target,
as compared to an otherwise in situ (passive) target.
We might also break AT into simple
and complex. In simple AT the approach is one of applying
heat and looking for a defining delta T on the surface of
a target. Complex active thermography generally employs one
or more requirements for time, intensity, sequence synching,
heat generation/application and temperature data/view in correlation
with observed Delta T’s on the target’s surface.
The additional features of complex AT allow its use in material
irregularities that otherwise might not be visible in simple
AT. The additional features may also provide more definition,
depth and site detail.
Discussion
Simple AT approaches supply quick (scan
time) and, when practiced, repeatable results. Examples of
simple approaches might include: alternate heated and cooled
rooms in a building to allow viewing possible moisture in
walls between rooms, supplying hot water to cold water pipes
to find water line leaks in building concrete slabs and heating
auto dashboard surfaces to detect voids.
Complex AT approaches are more costly in
time and resources but generally provide more detailed results.
Examples will be provided later in this paper.
An example of simple AT developing into complex
AT might be found in the military use of heating the surface
of a composite aircraft wing and looking for voids or delaminations.
The first approach of using hot air took time. Next, a pulse
of high intensity light was used to quickly heat and allow
for quick cooling of the surface. But this required faster
IR scan rates, which was offset by synching the flash and
image timing. Today, even the spectrum of light and timing
is varied to affect the heating or viewing of the composite
or metal surface.
History of
AT Development
Active thermography is an outgrowth of questions
about the possible use of newer and better IR equipment. When
infrared imagery was first applied to the military, civilian
and research sectors of industry, typical uses were found
for the new technology. Since its inception, infrared imagery
has been identified as a leader in obtaining quick answers
to non-intrusive testing questions. As more industries looked
for non-intrusive answers to normally destructive questions,
thermography was promoted even more. As more uses for thermography
were found, more ways to make the IR equipment perform were
found. We eventually started heating the targets to obtain
more pronounced or prolonged surface views.
Individuals like Dr. Edmund G. Henneke, have
detailed and pioneered such methods as vibrothermography techniques
since the late 1970’s. Each specialized technique has
individuals and organizations developing and making improvements
to their technique. Another example of challengers in their
specialized AT technique would be J. R. Lesniak and B. R.
Boyce in their work associated with thermal stress analysis
(TSA). The list of contributors to the advancement of AT techniques
is great and includes individuals, universities and research
organizations around the world.
Active thermography has a large repository
of history in development and use. Much more than could adequately
be covered in this paper and might well be a subject for later
detail.
Changing the Viewed
Target’s Surface Heat Pattern
Having made the decision that AT might provide
you with the needed testing, you now might also consider the
techniques for heating, cooling or stressing a target to cause
heating.
In performing AT, the types of thermal stress
are many and varied. Some of the more common would include
the use of:
Simple AT |
|
Complex AT |
| |
|
|
Complex AT |
|
Acoustics |
Air stream |
|
Induction heating |
Steam |
|
Pressure |
Current |
|
Microwaves |
Man-made light |
|
Torque |
Sunlight |
|
Man-made light |
When one or more of these stresses are applied
in conjunction with the complex AT methods, we derive some
of the specific techniques that are now in use or are being
developed and refined.
We should note at this point that there are
two basic methods of applying stresses to a target.
The first is transmissive and involves applying
the stress from behind the intended viewing area. This method
often is more time consuming in that viewed response is dependent
upon thermal characteristics, i.e., conduction and diffusion
and thickness of the target. This method also produces a saturated
(heat laden/loaded area) that must be allowed to cool before
areas of interest can again be viewed.
In contrast, the second method, the front
surface method, will typically produce meaningful data relatively
fast and may be controlled to keep from saturating the target
area thermally. In this method, stress is applied at the viewed
surface of a target and then removed to allow viewing with
IR instruments.
It is possible to combine these methods
in some forms of complex AT. An example might be the use of
a stacked ultrasonic transducer located at a corner of a test
specimen. The vibrations induced can be continuous and result
in flaw detection throughout the test specimen. Another example
might be the use of eddy current induction, from behind or
the side of the specimen, to enhance possible defects.
Types Of and Techniques
Employed in Complex AT
Traditionally there are two basic methods
of complex AT. They are pulsed and lock-in thermography.
Pulsed thermography employs short term heating
(pulses of heat) of a target and then monitoring the heat
decay as observed on the surface of the target. This is generally
a qualitative IR technique. In pulsed thermography, heat is
measured in its transient state. Pulsed thermography provides
some of the quickest meaningful data in complex AT.
Lock-in thermography employs a sinusoidal
heat stress to a target and derives information like depth
and type of fault by applying vectored heat signature data.
In lock-in thermography, heat is measured in its steady state.
An additional method was developed in an
attempt to bridge these two methods and it is called pulsed
phase thermography. In this form of thermography, a short
(in the order of 1 millisecond), high energy burst typically
from a flash lamp, is applied to a target surface. The frequency
of pulse with frequency of heating cycle is applied to fourier
transforms. Additional depth, detail of fault, and type of
fault may be realized using this combined technique.
There appear to be many different forms of
AT techniques both already developed and in experimentation.
All are generated by a specific need in a specific industry
and seem to grow in use and complexity as different industries
find other uses for existing techniques. With complexity,
additional tools and techniques to better define a flaw or
repeatable characteristic in a target are incorporated.
When approaching the selection of a
complex AT technique, consider the basic block of a complex
AT system, example shown below.
Typical Block for Lock-In
or Pulsed Thermography
In the example above, the system processing
unit (SPU) acts as the heart of the operation. The SPU syncs
the heating source to the viewing time of the imager and
records/transmits the data to a display device. The imager
may be an IR spot, line imager or imaging radiometer. You
may also note that the typical complex AT uses the front
surface method of stressing the target.
In most cases, stress application and thermal
pattern observation time, in conjunction with processing
the observed data, are the parameters that dictate the method
names given to the many forms of complex AT. Some of the
typical methods include:
1. Vibrothermography (VT)
a. Ultrasonic Lock-In Thermography (ULT)
b. Ultrasonic Burst Phase Lock-In Thermography
(UBP)
2. Thermoelastic Stress Analysis (TSA)
3. Optically Excited Lock-In Thermography
(OLT)
Vibrothermography
(VT)
Vibrothermography is an active IR technique
that allows viewing the surface of a specimen or target
and detecting problems in the surface and the bonding in
layers beneath or within laminated or multi-layered products.
When a sonic wave is used to vibrate
the surface of a solid or layered target, the sonics radiate
outward from the source. Inclusions, determs and disbonds,
and cracks will resonate at the frequencies of the induced
sonics and produce heated patterns on the surface of the
target. The thermoelastic effect (in metals) and the hysteresis
loss effect (in polymers) become pronounced with mechanical
vibration of materials at fault sites and result in resonance
at the fault site.

Vibrothermography
Operational Block
The rate of defect heating is proportional
to the frequency of the mechanical signal. This is why the
signal is usually ultrasonic. At this point, the defect
may be discerned as somewhere beneath the area shown by
this technique. By adding low frequency modulation and phase
locking to the ultrasonic signal (ULT or UBP), depth information
of the possible defect may also be obtained. This is due
in part to processing surface patterns as a vectored response
to the sonics. When ultrasonic bursts are passed through
a target, internal defects may cause parasitic (detracting
or canceling) resonant waves to occur, which in turn produces
a standing wave that may mask or detract from thermal patterns
representing the defect. By applying frequency modulated
ultrasound, these unwanted standing waves may be reduced
or eliminated. This will then allow better resolution and
measurement of suspected areas of concern.
Vibrothermography has current applications
in wood, polymers, ceramics, and metals. It should be noted
that it takes a lot of initial data via experimentation
with samples, and development of methods applied to a target
to produce repeatable and predictable results. Once initial
constraints and methodology are obtained, repeatable results
usually vary only with users of the technique and their
equipment.
Thermoelastic
Stress Analysis Thermography (TSA)
This is an active IR technique that allows
viewing thermal variations at the surface of an object or
structure by cyclic variation of a stress (pressure) applied
to the object.
When pressure is applied to a metal
block, the block will heat up. By applying the pressure
in a sinusoidal cycle to an elastic material, clear patterns
of heating are viewed by IR imagery across the surface of
the material. The cyclic pressure allows a cool down period
and also provides for viewing a second order harmonic, which
in turn allows for validation of the first order harmonics
viewed results.
Thermoelastic Stress
Analysis Operational Block
One current need for this form of
active thermography is in evaluation of materials / components
during fabrication and in situ usage. During manufacture
of some high stress components, pre-stressing of the surface
adds to the object’s ability to inhibit cracking.
With manufacturing processes and usage fatigue, these pre-stressed
components can weaken. This will lead to possible early
damage or failure of the component. Examples of materials
tested include Inconel 718 and Titanium-based alloys used
in aviation engines.
TSA is another form of active thermography
that uses electronic lock-in techniques to synch and coordinate
the process.
Optically
Heated Lock-In Thermography (OLT)
This method of AT refers to use of
man-made light sources to stimulate the target. The target
is heated while viewing time is coordinated with the cycle
of heating stimulus. This allows viewing the target while
target temperature is in an apparent steady state.
Optically Heated Lock-In
Thermography Block
Having viewed some of the types of and approaches
to AT you might consider how to decide which type might assist
you in your testing.
Current Applications
for AT
When considering typical applications for
AT, it might be beneficial to separate these uses into application
groups of simple or complex AT.
These groups include just a few of the many
possibilities available.
Simple AT would include inspections
of:
|
Composites |
|
for inclusions, disbonds and voids |
|
Cellulose products |
|
for entrapped moisture (fungi inspection) |
|
Painted surfaces |
|
for subsurface corrosion |
|
Bridge decks |
|
for girder separations |
Complex AT would include inspections of:
| Ultrasonic Stressing |
|
|
Polymers
|
|
for cracks, separation of laminates,
interaction between laminates and corrosion |
Metals
|
|
for cracks, weld flaws, deformation |
Ceramics
|
|
for cracks or bonding |
Cellulose
|
|
for separation of laminates |
| |
|
|
| Optical Stressing
|
|
|
Polymers
|
|
for voids, inclusions, and determ problems while possibly
providing size, type and depth |
| |
|
|
| Pressure Stressing |
|
|
Metals
|
|
for surface stress, tension or fracture
|
Ceramics
|
|
for tension and cracking |
| |
|
|
| |
|
|
One Approach for
Application of an AT Technique to Your Needs
Somewhere, someone wanted to know if
a separation of bridge decks from support girders (stress
or corrosion causal) could be observed and possibly classed.
Aircraft with newly developed composite wings were launched
off aircraft carriers, leaving layers of aircraft wings behind
and then landing with wings that were half their original
thickness. In this case, traditional use of ultrasound to
determine delamination did not work. Another industry wanted
to know if corrosion and its degree could be detected under
painted surfaces. Still another industry was looking for a
way to detect voids in foam panels as they moved down an assembly
line. The list of questions and applications for AT seems
endless and is bound only by whether the question has been
asked and subsequently researched.
So, you are now the one asking if AT might
provide a possible solution to your need. Perhaps the following
approach will help:
a. Determine what you hope to find in the
target when using AT
b. Understand the expected thermal properties
of the target’s surface and subsurface, considering
heat transfer through the target
c. Decide on the expected minimum size of
flaw or characteristics you hope to observe
d. Review the currently employed techniques
of AT and decide if one or more may provide you with the data
needed
e. Discuss your intended use and expectations
with others currently using the decided upon technique
f. Determine equipment needs for your AT
intended technique
g. Perform and record the results of your
AT. Records should be such that repeatable effort and data
are obtained. Baseline data will be required in most uses
of complex AT for use in your application.
Conclusion
In closing, it should be noted that this
paper was intended as a brief introduction to a continuously
expanding field of thermography. The fastest changes in AT
are occurring around its applications. That is where you come
in.
Our customers continue to ask us if our imagery
can be used to better their product or process. “Active
Thermography” may well be a near future answer from
you.
References
1. Nondestructive Testing Handbook, Volume
3, Third Edition, Infrared and Thermal Testing, ASTM ©
2001, Chapter 11, “Techniques of Infrared Thermography”
2. NASA/TM 2001-210697, “Quantifying
Residual Stresses by Means of Thermoelastic Stress Analysis”,
Andrew L. Geyenkenyesi (Ohio Aerospace Institute, Brook Park,
Ohio) and George Y. Baaklini (Glen Research Center, Cleveland,
Ohio) February, 2001
3. 16th ASCE Engineering Mechanics Conference,
July 16-18, 2003, University of Washington, Seattle, “A
Quantitative Thermoelastic Stress Analysis for Poltruded Composites”,
Rani F. El-Hajjar and Rami M. Haj-Ali
4. “Experimental Considerations in
Vibrothermography”, Steven M. Shepard, Tasdique Ahmed
and James R. Lota, Thermal Wave Imaging, 845 Livernois, Ferndale,
MI 48220
5. “Non-Destructive Testing of Composites”,
Ron Predmesky, IR/INFO 2002, Orlando, FL
6. “PHD Research Thesis: An Inverse
Problem Arising in Vibro-Thermography”, Thomas Hoft,
January 11, 2005
7. “Pulsed Phase Infrared Thermography”,
X. Maldaque, S. Marinetti, Electrical and Computing Engineering
Department, University Leval, Quebec City, P. Que., Canada
G1K 7P4
8. “Common Sense Approach to
Thermal Imaging”, Gerald C. Holst, Published by JCD
Publishing and SPIE Optical Engineering Press, Copyright 2000
|